Effect of Waterlase laser retrograde root-end cavity preparation PDF Ispis E-mail


Effect of Waterlase laser retrograde root-end cavity preparation on the integrity of root apices of extracted teeth as demonstrated by light microscopy

James A. Wallace, DDS, MDS, MSD, MS

School of Dental Medicine, University of Pittsburgh, Pittsburgh, Pennsylvania, USA

Keywords Abstract

...
 

apical preparations, root-end cracks, root-end preparatioans, Waterlase laser. Most endodontists use ultrasonic instruments for retrograde root-end cavity preparations even though they have been found to produce cracks. In this labo ratory study, thirty-six randomly chosen roots had root-end cavity preparations

Dr James A. Wallace, School of Dental Medicine,

made with the Waterlase laser and only one questionable intra-canal crack was

University of Pittsburgh, 3501 Terrace Street,

found. It was concluded that the Waterlase laser when used to make endodontic

3064 Salk Annex, Pittsburgh, PA 15261, USA.

root-end cavity preparations produces either no cracks, or a very low percentage

Email: Ova email adresa je zaštićena od spam robota, nije vidljiva ako ste isključili Javascript

(2.8%) of cracks.

doi: 10.1111/j.1747-4477.2006.00006.x

Introduction

Most endodontists consider ultrasonics as the method of choice for retrograde root-end cavity preparation. Several authors have found ultrasonic root-end cavity preparation produces cracks and/or chipping. (Table 1).

The Biolase Waterlase (Biolase ® Technology, Inc. San Clemente, CA, USA) Er-Cr:YSGG (Erbium, Chromium: Yttrium, Scandium, Gallium and Garnet) pulsed laser has been found to be useful in endodontic surgery for root-end resection, root-end cavity preparation, haemostasis, and sterilization of the root apex and surrounding tissue. For restorative procedures, laser use has increased patient acceptance related to pain, vibrations, whine of the drill, micro-fractures and heat production (15). This laser cuts hard tissue with highly energised water particles and soft tissue directly with laser energy. (15) Preliminary studies looking at the safety and efficacy of using the ErCr:YSGG laser found it to be a proficient instrument in cutting bone.

(16) FDA approval for apicoectomy surgery was granted on 12 February 2002 and flap surgery on 3 February 2003. Gouw-Soares (17) demonstrated the Er:YAG (Erbium:Yttrium-Aluminium Garnet), Ga-Al-As (Gallium Aluminium Arsenide) and Nd:YAG (Neodymium: Yttrium-Aluminium Garnet) lasers, when used in combination for performing an apicoectomy, produced heat which may lead to cracking. Root-end cavity preparation should be three millimetres in depth and the resection angle should be zero (18,19).

The purpose of this study was to determine if root-end preparations at a depth of three millimetres in resected roots at a zero angle performed by the Waterlase laser produce cracks and/or chipping. To the author’s knowledge there is no published data on the use of the Waterlase laser for this purpose.

Materials and methods

Seventeen extracted teeth, comprising eight mandibular molars, five maxillary molars, three bicuspids and one central incisor were chosen at random for a total of thirty-six root apices.

The teeth were stored in 0.9% sodium chloride and 1% sodium hypochlorite solution to preserve and inhibit microbial growth. All the teeth apices were preoperatively evaluated by two independent investigators with a fibre-optic translucent light source for a time period not exceeding 2 min using a Fisher stereomicroscope (FSM) at ×40 magnification and a Global Surgical Microscope (GSM) at ×12 magnification with digital photographs being taken at this time (20).

© 2006 The Authors Journal compilation © 2006 Australian Society of Endodontology

Effect of Waterlase Laser J. A. Wallace et al. Table 1 Ultrasonic and bur root-end preparation in vitro

Paper Type of instrument Frequency of cracks Assessment
Abedi et al. (1) Fissure bur and ultrasonic Significantly more cracks with ultrasonic versus bur SEM
Layton et al.(2) Ultrasonic low and high More than 40% demonstrated cracks Dye and microscope
Frank et al. (3) Bur slow and high speed, sonic 10–50% of teeth Dye and microscope

ultrasonic medium and high

Lloyd et al. (4) Sonic and bur Sonic 10–15% SEM


Bur 0–5%
Beling et al.(5) Ultrasonic 5–10% SEM
Min et al.(6) Bur, ultrasonic Bur 10%, ultrasonic 100% Confocal microscopy

Low and high

Brent et al. (7) Ultrasonic 20–25% SEM
Morgan and Marshall (8) Ultrasonic 5% SEM
Gray et al. (9) Bur and ultrasonic Bur 0% SEM


Ultrasonic 7%
Rainwater et al. (10) Bur and ultrasonic 60–80% ultrasonic Microscope


10% bur
Peters et al. (11) Ultrasonic 1% SEM
Gondim et al. (12) Sonic, ultrasonic 18–80% SEM
Ishikawa et al. (13) Ultrasonic 10–20% SEM
Khabbaz et al. (14) Bur, sonic, ultrasonic 7–20% Video microscope

SEM, standard error of the mean.

A three millimetre root-end resection was made on each root apex perpendicular to the long axis using a carbide bur in a high speed hand piece with water using GSM at ×12 magnification (21). The resected root apex was again examined for cracks and photographed as above. Following root resection, the teeth were immediately placed in a solution of 0.004% aqueous methylene blue dye in distilled water (2). Forty-eight hours following immersion, two investigators independently examined the specimens using the FSM ×40 magnification and photographed them with the GSM at ×12 magnification (2).

The root canals of the teeth used in the study were uninstrumented and unobturated. The root-end preparations were made using the Waterlase laser with a 600 µm laser tip and a setting of four watts, 55% water and 65% air as suggested by the manufacturer. The tip was used 1–2 mm from the surface using the GSM at ×12 magnification. A Class I root-end preparation was made in the 36 resected root-ends to a depth of 3 mm and approximately 1 mm in diameter confirmed with a periodontal probe. The root-end preparation was done by holding the teeth in the operator’s gloved hand with the tooth surrounded by saline-soaked gauze. The root was kept moist during preparation as energised water molecules did the cutting. Following apical preparation the specimens were examined and photographed with the GSM ×12 magnification (Fig. 1).

The teeth were immediately returned to the storage unit and immersed in 0.004% methylene blue solution. The total preparation time for each root was less than 2 min.

The teeth were then re-examined under the FSM at ×40 magnification with transillumination by two independent investigators.

Results

The 36 roots were evaluated before resection with the GSM ×12 magnification and the FSM at ×40 magnification by two independent investigators using transillumination for under 2 min and no cracks were identified. Digital photographs with GSM at ×12 magnification illustrated no cracks.

The root-ends were evaluated after resection as would be done in a clinical surgical situation, and no cracks were evident using the GSM at ×12 magnification. Forty-eight hours following immersion in 0.004% methylene blue dye, two investigators independently examined the resected root ends with the GSM ×12 magnification and the FSM at ×40 magnifications with transillumination, and once again no cracks were found.

During the root-end preparation in handling the specimens with gloved hands no detectable heat was produced in the specimens, thus enhancing chances for no cracks being produced. Sample specimens before and after root-end preparation are shown in Figures 1–3.

The teeth were re-examined under the FSM at ×40 magnification using transillumination by two independent investigators. One questionable intracanal crack was observed in one root-end preparation (that is, in 2.8% of specimens).

J. A. Wallace et al. Effect of Waterlase Laser

Discussion and conclusion

Pulsed ErCr: YSGG laser energy can be used to prepare root-ends for an apical seal. It is the energised water molecules that do most of the cutting and thus it was found that the roots remained very cool to the touch during preparation, as has been noted for osseous tissue (22). No cracks were noted pre-treatment. The digital photographs with the GSM at ×12 magnification were studied and no cracks were observed. It is important to note that this is the magnification typically used during clinical surgical procedures. Scanning electron microscopy examination would have been useful for closer inspection of the samples. The canals were left un-instrumented as there was no significant difference in the incidence of root cracks when canals were obturated or unobturated (6). Three millimetres of the root apices were removed with a high-speed fissure bur and water under the GSM at ×12 magnification to simulate clinical conditions and no cracks were observed. The root-end preparations were prepared to a depth of three millimetres with the Waterlase laser using laser settings recommended by the manufacturer. There were again no cracks observed during or immediately after the procedure.

The thickness of remaining dentine is not of such a concern with the laser preparation, as it would be with ultrasonic or rotary instruments, because there is no vibration or pressure exerted during root-end preparation that may produce cracks. Methylene blue plus transillumination with magnification was used to detect dentinal cracks as

© 2006 The Authors Journal compilation © 2006 Australian Society of Endodontology

Effect of Waterlase Laser J. A. Wallace et al.

recommended by Wright et al. (23). After 24 h in the methylene blue dye, the teeth were again evaluated and photographed with the GSM ×12 magnification. They were again subjected to transillumination and FSM at ×40 magnification and evaluated for under 2 min. Only one questionable intracanal crack was found. This may have initially been present but could not be detected until the root-end preparation was completed.

Connective tissue changes that occurs in response to other laser root surgery would not occur with the Water-lase as it is the energised water that does the cutting, not the laser. At present, no apical preparation laser micro-handpiece is available but such an instrument is under development. Teeth in situ, regardless of the method of root-end preparation, do not exhibit a lesser tendency towards cracking than extracted teeth (12).

Based on this laboratory study, the Waterlase laser does not produce a clinically relevant rate of cracking when used to make endodontic root-end preparations. The next step is to use this laser in root-end preparations under clinical conditions with the GSM and to record the results with digital photographs.

Acknowledgement

The author is grateful to John R. Skoner, DMD, MDS for his help in evaluating the samples.

...
References
  1. Abedi HR, Van Mierlo B, Wilder-Smith P, Torabinjad M. Effects of ultrasonic root cavity preparation on the root apex. Oral Surg Oral Med Oral Pathol 1995; 80: 207–213.
  2. Layton CA, Marshall JG, Morgan LA, Baumgartner JC. Evaluation of cracks associated with ultrasonic root end preparation. J Endod 1996; 22: 157–60.
    1. Frank RJ, Antrim DD, Bakland LK. Effect of retrograde cavity preparation on root apices. Endod Dent Traumatol 1996;
    2. 12: 100–3.
  3. Lloyd A, Gutmann J, Dummer P, Newcrombe R. Microleakage of Diaket and amalgam in root-end cavities prepared using Micromega sonic retro-prep tips. Int Endod J 1997; 30: 196–204.
  4. Beling K, Marshall J, Morgan L, Baumgartner T. Evaluation for cracks associated with ultrasonic root-end preparation of gutta-percha filled canals. J Endod 1997; 23: 723–6.
  5. Min MM, Brown CE, Legan JJ, Kofrawy AA. In vitro evaluation of effects of ultrasonic root-end preparation on resected root surfaces. J Endod 1997; 23: 624–8.
  6. Brent PD, Morgan LA, Marshall JG, Baumgartner JC. Evaluation of diamond-coated ultrasonic instrumentation for root-end preparation. J Endod 1999; 25: 672–5.
  7. Morgan LA, Marshall JG. A scanning election microscopic study of in vivo ultrasonic root-end preparations. J Endod 1999; 21: 567–70.
  8. Gray GS, Hatton S, Haltzman DJ, Jenkins DB, Nielsen CJ. Quality of root-end preparation using ultrasonic and rotary instrumentation in cadavers. J Endod 2000; 26: 281–83.
  9. Rainwater A, Jeansonne BG, Sarkaro N. Effects of ultrasonic root-end preparation on micro-crack formation and leakage. J Endod 2000; 26: 72–5.
  10. Peters CI, Peters OA, Barbakow F. An in vitro study comparing root-end cavities prepared by diamond-coated and stainless steel ultrasonic retrotips. Int Endod J 2001; 34: 142–48.
  11. Gondim E Jr, Fiquesredo A, Ferraz CC, Teixeira FB, de Souza-Filho FJ. Effect of sonic and ultrasonic retrograde cavity preparation on the integrity of root apices of freshly

J. A. Wallace et al.

extracted human teeth: scanning election microscopy analysis. J Endod 2002; 28: 646–50.

  1. Ishikawa H, Sawada N, Kaybayashi C, Suda H. Evaluation of root-end cavity preparation using ultrasonic retrotips. Int Endod J 2003; 36: 586–91.
  2. Khabbaz MG, Kerezaudia NP, Aroni E, Tsatsas V. Evaluation of different methods for the root-end cavity preparation. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 2004; 98: 273–82.
  3. Van As G. Erbium lasers in Dentistry. Dent Clin North Am 2004; 48: 1017–59.
  4. Eversole LR, Rizoiu IM, Kimmel A. Osseous repair subsequent to surgery with an erbium hydrokinetic laser system. Presented at the International Laser Congress. International Proceedings Division; 25–28 September 1996. Athens, Greece.
  5. Gouw-Soares S, Tangi E, Haypel P, Cardoso W, Edwardo CP. The use of Er: YAG, Nd: YAG, and Ga-Al-As lasers in periapical surgery: a three-year clinical study. J Clin Laser Med Surg 2001; 19: 193–8.

Effect of Waterlase Laser

18. Mattison GD, von Fraunshafer JA, Delavanis PD, Anderson AN. Microleakage of retrograde amalgams. J Endod 1985;

11: 340–5.

  1. Gagliani M, Taschieri S, Molinari R. Ultrasonic root-end ultrasonic root-end preparation: influence of cutting angle on the apical seal. J Endod 1998; 24: 726–80.
  2. Carr GB. Advanced technique and visual enhancement for endodontic surgery. Endod Rep 1992; 7: 6–9.
  3. Nedderman TA, Hartwell GR, Partell FR. A comparison of root surfaces following apical root resection with various burs, scanning electron microscopy evaluation. Endod J 1998; 14: 423–7.
  4. Glockner K, Rumpler J, Ebeleseder K, Stodler P. Intrapulpal temperature during preparations with the ER:YAG laser compared to the conventional bur: an in vitro study. J Clin Med Surg 1998; 16: 153–7.
  5. Wright HM, Loushine RJ, Weller RN, Kimbrough WF, Waller J, Pashley DH. Identification of resected root-end dental cracks. A comparative study of transillumination and dyes. J Endod 2004; 30: 712–15.

© 2006 The Authors Journal compilation © 2006 Australian Society of Endodontology


 